Edge Ultrasound System Guia do Utilizador Página 1

Consulte online ou descarregue Guia do Utilizador para não Edge Ultrasound System. Edge Ultrasound System User guide Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 3

88 Measurement publications and terminology

Página 4

Chapter 6: Troubleshooting and Maintenance 89TroubleshootingChapter 6: Troubleshootingand MaintenanceThis chapter contains information to help correc

Página 5 - Contents

90 Software licensing patient” Delete any unwanted images or clips from the patient exam. See “To delete images and clips” on page 41.Software licens

Página 6

Chapter 6: Troubleshooting and Maintenance 91TroubleshootingCleaning and disinfectingUse the recommendations in this section when cleaning or disinfe

Página 7

92 Cleaning and disinfecting To clean the LCD screen Dampen a clean, non-abrasive, cotton cloth with an ethanolic-based cleaner, and wipe the screen

Página 8

Chapter 6: Troubleshooting and Maintenance 93TroubleshootingTo clean and disinfect a transducer (wipe method)1 Disconnect the transducer from the sys

Página 9

94 Cleaning and disinfecting 2 Clean the surface using a soft cloth lightly dampened in a mild soap or detergent cleaning solution.Apply the solution

Página 10

Chapter 7: Safety 95SafetyChapter 7: SafetyThis chapter contains ergonomic, electrical, and clinical safety information required by regulatory agenci

Página 11 - Introduction

96 Position the systemPromote comfortable shoulder, arm, and hand postures• Use a stand to support the weight of the ultrasound system.Minimize eye a

Página 12

Chapter 7: Safety 97Safety• Minimize the pressure applied on the patient.• Keep your wrist in a straight position.Take breaks, exercise, and vary act

Página 13 - Chapter 1: Getting Started

ixIntroductionIntroductionEdge Ultrasound System User Guide provides information on preparing and using the Edge™ ultrasound system and on cleaning a

Página 14 - To remove the battery

98 WARNING:To avoid the risk of electrical shock: • This equipment must be connected only to a supply mains with protective earth. • Use only properl

Página 15 - Connecting transducers

Chapter 7: Safety 99SafetyElectrical safety classificationWARNING:To prevent injury to the operator/bystander, the transducer must be removed from pa

Página 16 - The system does not support

100 Equipment safetyTo protect your ultrasound system, transducer, and accessories, follow these precautions.Battery safetyTo prevent the battery fro

Página 17 - System controls

Chapter 7: Safety 101SafetyClinical safetyWARNING:Do not use a damaged battery.WARNING:Do not solder a battery.WARNING:The polarity of the battery te

Página 18 - 6 System controls

102 Hazardous materialsWARNING:Do not use the system if it exhibits erratic or inconsistent behavior. Discontinuities in the scanning sequence are in

Página 19 - Screen layout

Chapter 7: Safety 103SafetyElectromagnetic compatibilityThe ultrasound system has been tested and found to comply with the electromagnetic compatibil

Página 20 - General interaction

104 Electrostatic dischargeCaution:Electrostatic discharge (ESD), or static shock, is a naturally occurring phenomenon. ESD is common in conditions o

Página 21 - Annotation and text

Chapter 7: Safety 105SafetySeparation distanceCompatible accessories and peripheralsSonoSite has tested the Edge ultrasound system with the following

Página 22 - Preparing transducers

106 You may use these SonoSite accessories and third-party peripherals with the Edge ultrasound system. WARNING:Use of the accessories with medical s

Página 23 - Training videos

Chapter 7: Safety 107SafetyGuidance and manufacturer’s declarationThe Edge ultrasound system contains an IEEE 802.11 transmitter that utilizes the IS

Página 24 - Intended uses

x Customer comments

Página 25

108 Guidance and Manufacturer’s Declaration - Electromagnetic EmissionsThe Edge ultrasound system is intended for use in the electromagnetic environm

Página 26 - 14 Intended uses

Chapter 7: Safety 109SafetyElectrostatic Discharge (ESD)IEC 61000-4-2 Battery Power±6.0KV contact±8.0KV air±6.0KV contact-2KV air / +4KV airESD preca

Página 27 - Chapter 2: System Setup

110 Power Frequency Magnetic FieldIEC 61000-4-83 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location i

Página 28 - User setup

Chapter 7: Safety 111SafetyFCC Caution: Changes or modifications not expressly approved by the party responsible for compliance could void the user’s

Página 29 - Chapter 2: System Setup 17

112 Immunity testing requirementsThe Edge ultrasound system complies with the essential performance requirements specified in IEC 60601-1-2 and IEC 6

Página 30 - Annotations setup

Chapter 7: Safety 113SafetyBiological riskDevice complies with relevant Brazilian regulations for electro-medical devices.Canadian Standards Associat

Página 31 - Connectivity setup

114 Do not stack over 5 high.Do not stack over 10 high.Electrostatic sensitive devicesDevice complies with relevant FCC regulations for electronic de

Página 32 - OB Calculations setup

Chapter 7: Safety 115SafetyTemperature limitationAtmospheric pressure limitationHumidity limitationSubmersible. Protected against the effects of temp

Página 33 - OB Custom Tables setup

116 SpecificationsDimensionsSystem• Length: 13 in. (33 cm)• Width: 12.4 in. (31.5 cm)• Height: 2.5 in. (6.3 cm)Display• Length: 9.7 in. (24.6 cm)• He

Página 34 - USB Devices setup

Chapter 7: Safety 117SafetyShipping and storage (battery)-20–60°C (-4–140°F), 15–95% R.H. (For storage longer than 30 days, store at or below room te

Página 35 - Limitations of JPEG format

Chapter 1: Getting Started 1Getting StartedChapter 1: Getting StartedAbout the systemThe Edge ultrasound system is a portable, software-controlled de

Página 36 - 24 USB Devices setup

118 EMC standards classificationCISPR 11, International Electrotechnical Commission, International Special Committee on Radio Interference. Industria

Página 37 - Chapter 3: Imaging

Chapter 8: Acoustic Output 119Acoustic OutputChapter 8: Acoustic OutputThis chapter contains safety information required by regulatory agencies perta

Página 38 - M Mode imaging

120 The system has been designed to ensure that temperature at the face of the transducer will not exceed the limits established in Section 42 of EN

Página 39 - CPD and color Doppler imaging

Chapter 8: Acoustic Output 121Acoustic Output•Comet tailsFor more information on detecting and interpreting acoustic artifacts, see the following ref

Página 40 - PW and CW Doppler imaging

122 Table 2: TI (TIS, TIC, TIB) TransducerCPD SettingsPW SettingsBox WidthBox HeightBox DepthPRF Depth OptimizeC11x   (Depth)C60x   (PRF)HF

Página 41 - Spectral trace controls

Chapter 8: Acoustic Output 123Acoustic OutputOutput displayThe system meets the AIUM output display standard for MI and TI (See“Related guidance docu

Página 42 - Freezing, viewing frames

124 Even when MI is less than 1.0, the system provides a continuous real-time display of MI in all imaging modes, in increments of 0.1.The system mee

Página 43 - Needle visualization

Chapter 8: Acoustic Output 125Acoustic OutputThe displayed MI and TI values are based on calculations that use a set of acoustic output measurements

Página 44 - MBe subcontrols

126 Acoustic output measurementSince the initial use of diagnostic ultrasound, the possible human biological effects (bioeffects) from ultrasound exp

Página 45 - Imaging modes and exams

Chapter 8: Acoustic Output 127Acoustic OutputIn Situ= Water [e-(0.23alf)]where: In Situ = In Situ intensity valueWater = Water intensity value e = 2.

Página 46

2 Preparing the system To install the battery1 Disconnect the power supply from the ultrasound system.2 Remove the system from the mini-dock (if pres

Página 47 - Annotating images

128 interest is composed entirely of soft tissue. When the path contains significant amounts of fluid, as in many first and second-trimester pregnanc

Página 48 - Patient information form

Chapter 8: Acoustic Output 129Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 49 - To end the exam

130 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 50 - Images and clips

Chapter 8: Acoustic Output 131Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 51 - Reviewing patient exams

132 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 52

Chapter 8: Acoustic Output 133Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 53 - ECG Monitoring

134 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 54 - 42 ECG Monitoring

Chapter 8: Acoustic Output 135Acoustic Output(a) This index is not required for this operating mode; value is <1.(b) This transducer is not inten

Página 55 - Chapter 4: Measurements

136 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 56 - 2D measurements

Chapter 8: Acoustic Output 137Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 57 - Doppler measurements

Chapter 1: Getting Started 3Getting StartedTurning the system on or offTo turn the system on or off Press the power switch. (See “System controls” o

Página 58 - Automatic trace results

138 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 59 - General calculations

Chapter 8: Acoustic Output 139Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 60 - EMED calculations

140 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 61 - SET key

Chapter 8: Acoustic Output 141Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 62 - Volume calculations

142 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 63 - Volume flow calculations

Chapter 8: Acoustic Output 143Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 64 - To calculate volume flow

144 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 65 - Exam-based calculations

Chapter 8: Acoustic Output 145Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 66 - 54 Exam-based calculations

146 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 67 - Measurements

Chapter 8: Acoustic Output 147Acoustic Output(a) This index is not required for this operating mode; value is <1.(b) This transducer is not intend

Página 68 - To measure LVd and LVs

4 Preparing the system Inserting and removing USB storage devicesYou can use a USB storage device to import and export various logs and setup configu

Página 69 - To measure peak velocity

148 (a) This index is not required for this operating mode; value is <1.(b) This transducer is not intended for transcranial or neonatal cephalic

Página 70 - 58 Exam-based calculations

Chapter 8: Acoustic Output 149Acoustic Output(a) This index is not required for this operating mode; value is <1.(b) This transducer is not intend

Página 71

150 (a) This index is not required for this operating mode; value is <1.(b) This transducer is not intended for transcranial or neonatal cephalic

Página 72 - Cardiac Index (CI)

Chapter 8: Acoustic Output 151Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 73 - (TDI) waveform

152 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 74 - IMT calculations

Chapter 8: Acoustic Output 153Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 75

154 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 76 - To sketch IMT

Chapter 8: Acoustic Output 155Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 77 - OB calculations

156 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 78

Chapter 8: Acoustic Output 157Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 79

Chapter 1: Getting Started 5Getting StartedSystem controls123547812131914161815610911171 Control keys Adjust on-screen controls.2 Alphanumeric keys E

Página 80 - Small Parts calculations

158 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 81

Chapter 8: Acoustic Output 159Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 82 - Orbital calculation

160 ((a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic u

Página 83 - Vascular calculations

Chapter 8: Acoustic Output 161Acoustic Output(a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended

Página 84 - Patient report

162 (a)This index is not required for this operating mode; value is <1.(b)This transducer is not intended for transcranial or neonatal cephalic us

Página 85 - EMED and MSK worksheets

Chapter 8: Acoustic Output 163Acoustic OutputTerms used in the acoustic output tablesTable 40: Acoustic Output Terms and DefinitionsTerm DefinitionIS

Página 86 - To display an MSK worksheet

164 Acoustic measurement precision and uncertaintyAll table entries have been obtained at the same operating conditions that give rise to the maximum

Página 87 - Measurement accuracy

Chapter 8: Acoustic Output 165Acoustic OutputTable 41: Acoustic Measurement Precision and UncertaintyQuantityPrecision (% of standard deviation)Uncer

Página 89 - Measurement publications

167GlossaryGlossaryTermsFor ultrasound terms not included in this glossary, refer to Recommended Ultrasound Terminology, Second Edition, published in

Página 90

6 System controls 8SETCALIPERSets a trace measurement.Displays calipers on-screen for measuring.9 CALCS Turns the calculations menu on and off.10 Tou

Página 91

168 SonoHD2™ Imaging TechnologyA subset of the 2D imaging mode in which the 2D image is enhanced by reducing speckle noise artifact at tissue margins

Página 92

169GlossaryAbbreviationsAbbreviations in User InterfaceAbbreviation Definition+/× “+” Caliper/”×” Caliper RatioA“A” Wave Peak VelocityA PG “A” Wave P

Página 93 - Obstetrical references

170 Ao AortaAoD Aortic Root DiameterApical Apical ViewAPTD Anteroposterior Trunk DiameterAT Acceleration (Deceleration) TimeAUA Average Ultrasound Ag

Página 94 - Gestational age tables

171GlossaryCW Continuous Wave DopplerCxLen Cervix LengthD DiameterD Apical Distance ApicalDCCA Distal Common Carotid ArteryDECA Distal External Carot

Página 95 - References

172 EFW Estimated Fetal WeightCalculated from the measurements performed during the exam. The measurements used to determine EFW are defined by the c

Página 96 - Growth analysis tables

173GlossaryHR Heart RateICA Internal Carotid ArteryIMT Intima Media ThicknessIVRT Iso Volumic Relaxation TimeIVS Interventricular SeptumIVSd Interven

Página 97 - General references

174 LVDFS Left Ventricular Dimension Fractional ShorteningLVDs Left Ventricular Dimension SystolicLVEDV Left Ventricular End Diastolic VolumeLVESV Le

Página 98

175GlossaryMsk MusculoskeletalMV Mitral ValveMV Area Mitral Valve AreaMV Regurgitant FractionMitral Valve Regurgitant FractionMV Regurgitant VolumeMi

Página 99

176 PGmax Maximum Pressure GradientPGmean Mean Pressure GradientPGr Pressure GradientPHT Pressure Half TimePI Pulsatility IndexPICA Proximal Internal

Página 100

177GlossaryRVW Right Ventricular Free WallRVWd Right Ventricular Free Wall DiastolicRVWs Right Ventricular Free Wall SystolicS SonoHD technologyS/D S

Página 101 - Chapter 6: Troubleshooting

Chapter 1: Getting Started 7Getting StartedScreen layout1 Mode Data Area Current imaging mode information (for example, Gen, Res, THI, and PW).2 Orie

Página 102 - Maintenance

178 UA Ultrasound AgeCalculated on the mean measurements taken for a particular fetal biometry.Umb A Umbilical ArteryVA Vertebral ArteryVArty Vertebr

Página 103 - Cleaning and disinfecting

Index 179IndexIndexSymbols+/x measurement 45Numerics2D imaging 252D options 25AA & B shortcut keys 15abbreviations 169abdominal, intended uses 12

Página 104 - To clean the LCD screen

180 Index MV/AV area 57overview 53PHT 58PISA 54RVSP 58setup 19SV 60TDI 61VTI 57cardiac index (CI) 60cardiac output (CO) 60, 61cardiac references 77c

Página 105 - Caution:

Index 181Indexuser accounts 17Ffar 5fetal heart rate (FHR) 67flow sensitivity 27focal zones, optimize 25footswitch setup 15forms 6freeze 30Ggainadjus

Página 106 - Cleaning the footswitch

182 Index MBe. See SonoMBemeasurements+/x Ratio, Doppler 45See also calculations2D 44about 43Acceleration, Doppler 45accuracy 43, 75area, 2D 44autom

Página 107 - Chapter 7: Safety

Index 183IndexPW Doppler. See pulsed wave (PW) Doppler imagingRreferencescardiac 77general 85gestational age tables 82growth analysis tables 84obstet

Página 108 - Position yourself

184 Index user guide, conventions used ixuser setup 16uses, intended 12–13Vvariance 28vascularcalculations 71intended uses 13velocity measurement 45

Página 110 - WARNING:

P15200-01*P15200-01*

Página 112 - Battery safety

8 General interaction General interactionTouchpad and cursorUse the touchpad to adjust and move objects on-screen. The touchpad controls caliper posi

Página 113 - Clinical safety

Chapter 1: Getting Started 9Getting StartedAnnotation and textAlphanumeric keyboard1 TAB Moves cursor among fields in the forms, and tabs between tex

Página 114 - Hazardous materials

10 Preparing transducers SymbolsYou can enter symbols and special characters in select fields and forms. The symbols and special characters available

Página 115 - Electromagnetic compatibility

Chapter 1: Getting Started 11Getting StartedTo apply a transducer sheathSonoSite recommends the use of market-cleared, transducer sheaths for intraca

Página 116 - Electrostatic discharge

12 Intended uses Intended usesThe system is used with a transducer attached and is powered either by battery or by AC electrical power. The clinician

Página 117 - Separation distance

Chapter 1: Getting Started 13Getting StartedSuperficial Imaging Applications You can assess the breast, thyroid, testicle, lymph nodes, hernias, musc

Página 118

14 Intended uses

Página 119

Chapter 2: System Setup 15SetupChapter 2: System SetupThe system setup pages let you customize the system and set preferences.Displaying the setup pa

Página 120

16 Administration setup Security settingsSecurity settings on the system allow you to meet the applicable security requirements listed in the HIPAA s

Página 121

Chapter 2: System Setup 17Setup3 Under User Information, make changes as desired.4 Select Save.Any change to the user name replaces the previous name

Página 122

USER GUIDE

Página 123

18 Annotations setup Logging in as userIf user login is required, the User Login screen appears when you turn on the system. (See “To require user lo

Página 124 - Labeling symbols

Chapter 2: System Setup 19SetupTo import predefined label groups1 Insert the USB storage device that contains the label groups.2 On the Annotations s

Página 125 - Chapter 7: Safety 113

20 Date and Time setup Date and Time setupTo set the date and time On the Date and Time setup page, do the following:•In the Date box, type the curr

Página 126 - STERILE EO

Chapter 2: System Setup 21Setup3 Select the USB storage device, and select Export.All user-defined tables and measurements are copied to the USB stor

Página 127 - Chapter 7: Safety 115

22 Presets setup 2 Select the desired table (Gestational Age or Growth Analysis).3 In the Measurement list, select the measurement for the custom tab

Página 128 - Specifications

Chapter 2: System Setup 23Setup• SiteLink organizes files in a SiteLink-style folder structure. Clips export in H.264 video saved as MP4 files. To vi

Página 129 - Standards

24 USB Devices setup

Página 130

Chapter 3: Imaging 25ImagingChapter 3: ImagingImaging modesThe system has a high-performance display and advanced image-optimization technology that

Página 131 - Chapter 8: Acoustic Output

26 Imaging modes M Mode imagingMotion mode (M Mode) is an extension of 2D. It provides a trace of the 2D image displayed over time. A single beam of

Página 132 - Acoustic artifacts

Chapter 3: Imaging 27Imaging3 Press the MMODE key.The time scale above the trace has small marks at 200ms intervals and large marks at one-second int

Página 133 - Acoustic Output

ii SonoSite, Inc.21919 30th Drive SEBothell, WA 98021USAT: 1-888-482-9449 or 1-425-951-1200F: 1-425-951-1201SonoSite LtdAlexander House40A Wilbury

Página 134

28 Imaging modes PW and CW Doppler imagingPulsed wave (PW) Doppler and continuous wave (CW) Doppler imaging modes are optional features. PW Doppler i

Página 135 - Output display

Chapter 3: Imaging 29ImagingPW Doppler controlsIn PW Doppler imaging, you can set the following on-screen controls.Spectral trace controlsIn spectral

Página 136 - AIUM Medical Ultrasound

30 Adjusting depth and gain Adjusting depth and gainTo adjust depthYou can adjust the depth in all imaging modes but the trace modes. The vertical de

Página 137 - Related guidance documents

Chapter 3: Imaging 31ImagingNeedle visualizationAbout MBeThe MBe control turns on SonoMBe™ imaging, which enhances linear structures within a selecte

Página 138 - Acoustic output measurement

32 Needle visualization Figure 2 With a curved array transducer, only segments of the needle shaft may show:1 - Upper needle shaft2 - Unshown segment

Página 139 - -(0.069lf)

Chapter 3: Imaging 33Imaging• Back returns to the previous screen. If MBe is on, MBe is highlighted green and MBe appears in the mode data area. Pres

Página 140 - Acoustic output tables

34 Imaging modes and exams available by transducer Imaging modes and exams available by transducerImaging ModeTransducerExam Type12D2M ModeCPD3Color3

Página 141 - CPD/Color

Chapter 3: Imaging 35ImagingAnnotating imagesYou can annotate live images as well as frozen images. (You cannot annotate a saved image.) You can plac

Página 142 - PW Doppler

36 Patient information form To turn off text entry, press the TEXT key.To reset the home position1 Press the TEXT key.2 Using the touchpad or arrow k

Página 143 - Operating Mode:

Chapter 3: Imaging 37Imaging• Done to save changes and return to imaging.To end the exam1 Make sure that you have saved images and other data you wan

Página 144 - Table 9: Transducer Model:

iiiContentsIntroductionConventions, symbols, and terms ...ixCustomer comme

Página 145

38 Images and clips Images and clipsSaving images and clipsWhen you save an image or clip, it saves to internal storage. The system beeps afterward i

Página 146 - CW Doppler

Chapter 3: Imaging 39ImagingReviewing patient examsThe patient list organizes saved images and clips in patient exams. You can delete, view, print, o

Página 147 - HFL38x/13-6

40 Images and clips To review images and clipsYou can review images and clips in only one patient exam at a time.1 In the patient list, highlight the

Página 148

Chapter 3: Imaging 41Imaging5 Select Export.The files are finished exporting approximately five seconds after the USB animation stops. Removing the U

Página 149

42 ECG Monitoring ECG Monitoring controlsShow/Hide Turns on and off ECG trace.Gain Increases or decreases ECG gain. Settings are 0-20.Position Sets t

Página 150 - Table 15: Transducer Model:

Chapter 4: Measurements and Calculations 43MeasurementsChapter 4: Measurementsand CalculationsYou can measure for quick reference, or you can measure

Página 151 - Table 16: Transducer Model:

44 Measurements Note: Trace measurements cannot be edited once set. To improve precision of caliper placement  Do any of the following:• Adjust the

Página 152

Chapter 4: Measurements and Calculations 45Measurements4 Press the SELECT key.5 Using the touchpad, complete the trace, and press the SET key.See “To

Página 153

46 Measurements 3 Press the SELECT key.A second caliper appears.4 Using the touchpad, position the second caliper at the end diastole on the waveform

Página 154

Chapter 4: Measurements and Calculations 47Measurements• Pulsatility Index (PI)• End Diastolic Velocity (EDV)• Acceleration Time (AT)• Resistive Inde

Página 155 - Operating Mode:

iv OB Calculations setup ...20OB Custom Measureme

Página 156

48 General calculations Displaying, repeating, and deletingsaved measurements in calculationsTo display a saved measurement Do one of the following:

Página 157

Chapter 4: Measurements and Calculations 49MeasurementsFigure 2 Percent area reduction calculation of right carotid bulb To calculate percent area re

Página 158

50 General calculations Volume calculationsTo calculate volumeThe volume calculation involves three 2D distance measurements: D1, D2, and D3. After a

Página 159 - L38xi/10-5

Chapter 4: Measurements and Calculations 51MeasurementsVolume flow calculationsThe following table shows the measurements required to complete the vo

Página 160

52 General calculations The system is limited to the following sample volume sizes:• C11x transducer: 1, 2, 3 Gate Size (mm)• C60x and P10x transduce

Página 161 - CPD/Color

Chapter 4: Measurements and Calculations 53MeasurementsExam-based calculationsIn addition to the general calculations, there are calculations specifi

Página 162 - PW Doppler

54 Exam-based calculations LV…LVd RVW (M Mode)RVD (M Mode)IVS (M Mode)LVD (M Mode)LVPW (M Mode)COEF SVLVESVLVEDVIVSFTLVPWFTLVDFSCISILV Mass…LVs RVW

Página 163 - Table 28: Transducer Model:

Chapter 4: Measurements and Calculations 55MeasurementsP. Vein A (Doppler) VMaxAdur (Doppler) timeS (Doppler) VMaxS/D ratioD (Doppler)MV E (Doppler)A

Página 164 - Table 29: Transducer Model:

56 Exam-based calculations To measure LVd and LVs 1 On a frozen 2D image or M Mode trace, press the CALCS key. 2 From the calculations menu, select t

Página 165

Chapter 4: Measurements and Calculations 57Measurementsc Using the touchpad, trace the left ventricular (LV) cavity.To make a correction, select Undo

Página 166

vVolume flow calculations ...51Exam-based calculations ...

Página 167 - Table 32: Transducer Model:

58 Exam-based calculations 4 Using the touchpad, trace the waveform.To make a correction, select Undo on-screen, backtrack with the touchpad, or pres

Página 168 - Table 33: Transducer Model:

Chapter 4: Measurements and Calculations 59MeasurementsTo make a correction, select Undo on-screen, backtrack with the touchpad, or press the BACKSPA

Página 169

60 Exam-based calculations a From the calculations menu, locate Qp/Qs and then select LVOT D or RVOT D.b Position the calipers. (See “Working with ca

Página 170

Chapter 4: Measurements and Calculations 61Measurements1 (CI Only) Fill in the Height and Weight fields on the patient information form. The BSA is c

Página 171

62 Exam-based calculations c Save the calculation. (See “To save a calculation” on page 47.)Gynecology (Gyn) calculationsGynecology (Gyn) calculation

Página 172

Chapter 4: Measurements and Calculations 63MeasurementsThe following table shows available measurements for IMT calculations. The IMT measurement nam

Página 173

64 Exam-based calculations IMT tool controlsWhen using the IMT tool, you can select the following controls on-screen.To trace IMT manuallyIn manually

Página 174

Chapter 4: Measurements and Calculations 65Measurementsa Position the caliper at the beginning of the boundary and press the SELECT key.b Using the t

Página 175

66 Exam-based calculations Results from System-Defined OB Measurements and Table Authorsa. The Gestational Age is automatically calculated and displ

Página 176 - 4 WoITA z

Chapter 4: Measurements and Calculations 67Measurementsauthor for the OB tables, which you choose on a system setup page, determines the measurements

Página 177

vi Equipment safety ...100Battery safety ..

Página 178

68 Exam-based calculations b Position the calipers:•For S/D, RI, position the first caliper at the peak systolic waveform. Press the SELECT key, and

Página 179 - Glossary

Chapter 4: Measurements and Calculations 69MeasurementsTranscranial Doppler and Orbital calculationsThe following table shows the measurements requir

Página 180

70 Exam-based calculations To perform a Transcranial Doppler or Orbital calculation1 Select the correct exam type:• Orbital (Orb) to measure Opthalmi

Página 181 - Abbreviations

Chapter 4: Measurements and Calculations 71MeasurementsVascular calculationsThe vascular measurements that you can save to the patient report are lis

Página 182

72 Patient report 3 Do the following for each measurement you want to take:a From the calculations menu, select the measurement name.b Using the touc

Página 183

Chapter 4: Measurements and Calculations 73MeasurementsTCD patient reportThe maximum values for the TAP calculation appear on the summary page.To del

Página 184

74 Patient report To display an MSK worksheetThe MSK worksheets have lists from which you can select and a field for entering comments.1 After or dur

Página 185

Chapter 5: Measurement References 75ReferencesChapter 5: Measurement ReferencesMeasurement accuracyThe measurements provided by the system do not def

Página 186

76 Sources of measurement errors Sources of measurement errorsIn general, two types of errors can be introduced into the measurement:Acquisition Erro

Página 187

Chapter 5: Measurement References 77Referencesfloating-point versus integer-type math, which is subject to errors introduced by rounding versus trunc

Página 188

viiGlossaryTerms ...

Página 189

78 Measurement publications and terminology Delta Pressure: Delta Time (dP:dT) in mmHg/sOtto, C.M. Textbook of Clinical Echocardiography. 2nd ed., W.

Página 190

Chapter 5: Measurement References 79ReferencesLVEDV = (7.0 * LVDD3)/(2.4 + LVDD)where: LVEDV = Left Ventricular End Diastolic VolumeLVDD = Left Ventr

Página 191 - Numerics

80 Measurement publications and terminology Mean Velocity (Vmean) in cm/sVmean = mean velocityMitral Valve Area (MVA) in cm2Reynolds, Terry. The Echo

Página 192 - See also

Chapter 5: Measurement References 81ReferencesRight Ventricular Systolic Pressure (RVSP) in mmHgReynolds, Terry. The Echocardiographer’s Pocket Refer

Página 193

82 Measurement publications and terminology EDD = LMP date + 280 daysEstimated Fetal Weight (EFW)Hadlock, F., et al. “Estimation of Fetal Weight with

Página 194

Chapter 5: Measurement References 83ReferencesHadlock, F., et al. “Estimating Fetal Age: Computer-Assisted Analysis of Multiple Fetal Growth Paramete

Página 195

84 Measurement publications and terminology Hadlock, F., et al. “Estimating Fetal Age: Computer-Assisted Analysis of Multiple Fetal Growth Parameters

Página 196

Chapter 5: Measurement References 85ReferencesObstetrics and Gynecology. Springer-Verlag, New York, (1986), 186, Table 7.20.)Femur Length (FL)Chitty,

Página 197

86 Measurement publications and terminology Acceleration Index (ACC)Zwiebel, W.J. Introduction to Vascular Ultrasonography, 4th ed., W.B. Saunders Co

Página 198 - *P15200-01*

Chapter 5: Measurement References 87ReferencesPulsatility Index (PI)Kurtz, A.B., W.D. Middleton. Ultrasound-the Requisites. Mosby Year Book, Inc., (1