Edge Ultrasound System Guia do Utilizador Página 151

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 150
Chapter 8: Acoustic Output 139
Acoustic Output
(a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
Data are not applicable for this transducer/mode.
Table 16: Transducer Model:
HFL50x
Operating Mode:
MMode
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.2 (a) (a) (b)
Associated Acoustic
Parameter
p
r.3
(MPa) 3.14
W
0
(mW) —# ##
min of
[W
.3
(z
1
),I
TA.3
(z
1
)]
(mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm) 1.4 #
d
eq
(z
sp
) (cm) #
f
c
(MHz) 6.75 # # #
Dim of A
aprt
X(cm) —# # #
Y(cm)
—# # #
Other Information
PD (μsec) 0.263
PRF (Hz) 1600
p
r
@PII
max
(MPa) 4.35
d
eq
@Pll
max
(cm) #
Focal Length FL
x
(cm) —# #
FL
y
(cm) —# #
I
PA.3
@MI
max
(W/cm
2
)
388
Operating
Control
Conditions
Control 1: Exam Type Any
Control 2: Optimization Pen
Control 3: Depth 4.0
Vista de página 150
1 2 ... 146 147 148 149 150 151 152 153 154 155 156 ... 197 198

Comentários a estes Manuais

Sem comentários